Features
- Fast response and high accuracy
- Long-term stable data for deployments of a year or longer
- Resistant to biofouling
- Real-time data output
- Large selection of concentration ranges
- Compact benchtop design
- Internal data logger and controller with 2GB flash memory
pCO2 Sensor Applications
- Ocean acidification studies
- Long-term pCO2 monitoring
- Shipboard flow-through pCO2 measurements
- Laboratory experiments
CO2-Pro™ FT pCO2 Sensor
The CO2-Pro™ FT instrument measures the partial pressure of CO2 gas dissolved in water using infrared detection. Standard ranges from 0-600 ppm up to 0 - 2,000 ppm, along with custom ranges, provide the full spectrum of pCO2 needed for accurate measurement of water CO2 levels.
Labelled FT for Flow-Through version of our flagship CO2-Pro™ instrument, the Pro™-FT can be used wherever a supply of flowing sample water is available, including shipboard flow-through systems and laboratory testing.
The small size of the CO2-Pro™ FT makes it versatile and easy to integrate into larger systems where space is a concern. When combined with our anti-fouling features and options, the instrument can perform in even the most biologically active waters for extended periods of time. An internal zeroing feature provides a stable long-term baseline to ensure accurate measurements.
The CO2-Pro™ FT is factory calibrated using WMO traceable standard gases. In addition, detector temperature stabilization and measurement of gas steam pressure and humidity provide an accurate and stable measurement of pCO2 in water.
PRODUCT SPECIFICATIONS
Sensor Performance | |
---|---|
Accuracy | ±0.5% |
Resolution | 0.01 ppm |
Zero drift | automatic zero compensation |
Equilibration time (t63) | 50 seconds (with water pumped head at 20º C) |
Standard range (alternate ranges available) |
0-600 ppm 0 - 1000 ppm 0 - 2000 ppm |
Physical | |
---|---|
Length | 20 cm (8 in) |
Width | 20 cm (8 in) |
Height | 10 cm (4 in) |
Weight (plastic) | 2.7 kg (6 lbs) |
Housing | Plastic or Aluminum |
Water Pressure | Up to 3 bar |
Water Temperature | 0º to 30º C (Standard) -2º to 20º C (Arctic) 15º to 40º C (Tropical) |
Electrical | |
---|---|
Input voltage | 10-18 VDC |
Power consumption | 3 W (9 W at warmup) |
Optional Water Pump | add 1 W |
Data output | RS-232, ASCII format, 0-5 V or 4-20mA optional |
Sample rate | 1 second (variable rate with logger/controller) |
OPTIONAL ACCESSORIES
Pigtail Cables with Locking Sleeve
5, 10, 25, 50 meters, or longer